99
50 – 250 μL ( Fig. 5)
Push the control button all the way down and pull off the
ejector sleeve (13).
Piston seal (4)
– Unscrew pipette at the central junction.
– Discard piston seal (4). Insert new piston seal into the
clamp (3) and place both into the lower part (5).
Ejector seal (10)
– Unscrew lower part of the pipette.
– Discard seal (10) with O-ring.
– Insert new seal (see "Inserting the seals").
Tighten the tip cone (see IV of this part).
The tip cone (11) is also available complete with seal.